Recall of Allura Xper Allura software versions: 8.1.25; 8.1.25.1; 8.1.25.5; 8.2.25; 8.2.25.5; and 8.2.27

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Philips Electronics Australia Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2018-RN-00796-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2018-06-20
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Philips have identified that the first time an operator selects a new procedure type during a single examination, the shutter position resets to the open position for the new procedure type. if the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.
  • Action
    Philips is advising this issue will be resolved in a software update, expected to be available in the second half of 2018. Philips will notify users when this software update is available. Until a software revision that corrects this issue becomes available, users should verify that the desired shutter position is set when performing the run after the first time the procedure type is changed during a single examination. This can be accomplished by first selecting a different procedure and then reselect the original procedure on the Xper Module or on the Data Monitor.