Recall of Alere Cholestech LDX Multianalyte Control (An in vitro diagnostic medical device used to monitor the performance of total cholesterol (TC), high density lipoprotein (HDL), triglycerides (TRG), glucose (GLU), alanine aminotransferase (ALT), and aspartate aminotransferase (AST) test procedures on the Alere Cholestech LDX System)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Inverness Medical Innovations Australia Pty Ltd T/A Alere.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-00227-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-03-14
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Investigations of the affected lot numbers have indicated that these lots of controls may yield results which are out of range high when tested with the alere cholestech ldx hdl test.
  • Action
    Inverness Medical Innovations (trading as Alere) is asking customers to discontinue the use of the affected lots and destroy all remaining stock. Customers have been advised that If the controls have been within the ranges shown on the value assignment card that came with the control materials, then the patient results should be considered clinically acceptable. Replacement product is expected to be available in approximately 4 – 6 weeks.

Device

  • Model / Serial
    Alere Cholestech LDX Multianalyte Control (An in vitro diagnostic medical device used to monitor the performance of total cholesterol (TC), high density lipoprotein (HDL), triglycerides (TRG), glucose (GLU), alanine aminotransferase (ALT), and aspartate aminotransferase (AST) test procedures on the Alere Cholestech LDX System)Catalogue Number: CT88769Lot Numbers: C2572, C2572A, C2593 & C2593AARTG Number: 160951
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    DHTGA