Recall of AIRO Mobile CT Scanner Model: MobiCT-32

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Brainlab Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-01256-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2017-09-29
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The manufacturer, mobius imaging, believes that due to an issue with the configuration settings of the tilt-drive motion controller, if a tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended tilt motion of gantry while airo system is homing the axis at startup or preparing for a scan. since the airo is not currently monitoring for intermittent connections or broken tilt-drive motor wires, if a wire breaks or the signal becomes intermittent the tilt-drive motor would not get the correct feedback and the motor could behave unpredictably. the following behaviours could be seen in the airo: when pushing the green button to home/tilt the gantry, the gantry may tilt in either direction, and at a higher than normal speed. the operator may not be able to respond (by releasing green button or pressing the emergency stop button) in time to prevent harm. to date, there have been no injuries reported related to this issue.
  • Action
    Due to the risk of unintended (tilting) motion of the gantry, customers are advised to discontinue using the AIRO until the tilt drive motion firmware has been updated on their system. Brainlab is arranging the firmware update for installation on the affected serial numbers on 29 September 2017.

Device

  • Model / Serial
    AIRO Mobile CT Scanner Model: MobiCT-32Serial Numbers: AIRO-0162, AIRO-0143, AIRO-0154, AIRO-0146, AIRO-0149, AIRO-0209 and AIRO-0217 ARTG Number: 223542
  • Product Classification
  • Manufacturer

Manufacturer