Recall of Aeroneb Pro/Solo Nebuliser Systems FRIWO power adapter

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Tag Medical Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-01003-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2014-09-18
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Tag medical have been advised by aerogen ltd that there is a potential hazard with certain friwo power adaptors as supplied with the aerogen pro and aerogen solo nebuliser systems manufactured during the period of week 50/2013 and week 23/2014. the power adapter could potentially fail, whereby the plastic screw pillars can break and dislodge from the body and the adapter cover can open.
  • Action
    Tag Medical is contacting the hospitals and replacing the faulty FRIWO power adaptors free of charge. This action has been closed-out on 06/06/2016.

Device

  • Model / Serial
    Aeroneb Pro/Solo Nebuliser Systems FRIWO power adapter Multiple catalogue numbers and date codes affectedARTG Number: 99080