Recall of Aequalis IM Nail Targeting Jig(Prosthetic joint implantation kit)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Tornier Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2015-RN-00587-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2015-07-07
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The targeting jig mast is separating from the jig boom. any separation will impact alignment and can affect the ability to fixate the screws.There has been no impact to a patient to-date; however, a design change to the product is necessary.
  • Action
    Tornier is developing a design change for the jig. Once the design change has been completed Tornier will arrange for the collection and replacement of the faulty devices. In the interim, Tornier has provided inspection instructions to ensure that faulty devices are not used in surgery.

Device

  • Model / Serial
    Aequalis IM Nail Targeting Jig(Prosthetic joint implantation kit)Lots: All lots - (DB0114069, DB0114349, CA46096, CA216052, DB0111264001, DB0112016, DB0113102)Catalogue Number: 9020060ARTG number: 177214
  • Manufacturer