Recall of Advantx, Innova 2000, Innova 2100 IQ, Innova 3100, Innova 3100 IQ, Innova 4100, Innova 4100 IQ Cardiovascular X-ray imaging systems equipped with Video Monitor Suspension (Fluoroscopic, angiographic x-ray system)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by GE Healthcare Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-00899-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-08-23
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    It has been identified during preventive maintenance activities that the 4 outer screws that hold the overhead video monitor suspension to the carriage were found missing since installation of the system. there is a potential that the system has not been installed per specified installation requirements. the missing screws may increase the risk for fall of the monitor suspension.
  • Action
    GE Healthcare is inspecting all potentially affected systems to verify that the overhead Video Monitor Suspensions are properly installed and will correct all affected systems.

Device

  • Model / Serial
    Advantx, Innova 2000, Innova 2100 IQ, Innova 3100, Innova 3100 IQ, Innova 4100, Innova 4100 IQ Cardiovascular X-ray imaging systems equipped with Video Monitor Suspension (Fluoroscopic, angiographic x-ray system)ARTG Number: 93871