Recall of ACE Control. An in vitro diagnostic medical device (IVD)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Immuno Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00156-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2014-02-14
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Upon review of the qa release testing, the range calculated at the time of kit release (done per standard operating procedures) supports the control range assigned upon product release, i.E. 28-52 at 37°c. however, upon investigation of a complaint encompassing several lots of reagents, the assigned control range has been reevaluated and it was determined that the range for these lots requires an adjustment to 23-43 at 37°c.This recall action was carried out prior to approval of the recall strategy by the therapeutic goods administration.
  • Action
    Immuno provided updated product information with the correct control ranges.

Device

  • Model / Serial
    ACE Control. An in vitro diagnostic medical device (IVD)Product Code: A6040Lot Numbers: 042, 043 and 044ARTG Number: 203619
  • Manufacturer

Manufacturer

  • Source
    DHTGA