Recall of ABL90 FLEX Analyser. An in vitro diagnostic medical device (IVD) (blood gas analyser)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Radiometer Pacific Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-01128-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2013-11-06
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Radiometer has recently become aware of a potential issue of negative bias on sodium results if the abl90 is not on a horizontal table. when tilted by 10 degrees or more and idle for 30 -45 minutes the magnitude of the effect on sodium results is as large as 10 mm (7%) in the worst case.
  • Action
    Radiometer is advising users to ensure that the ABL90 is placed on a horizontal platform/table. If the analyser is on a roll stand users must ensure that the support plate is aligned horizontally and locked in position. Radiometer will perform alignment of roll stands on request. A detection mechanism will be provided in an upgraded version of the analyser software to eliminate the possibility of error.

Device

  • Model / Serial
    ABL90 FLEX Analyser. An in vitro diagnostic medical device (IVD) (blood gas analyser)