Recall of 1.5T Signa HDxt, 3.0T Signa HDxt, Signa Vibrant MR Systems (Magnetic Resonance Imaging Systems)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by GE Healthcare Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-00454-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-05-16
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    For clariview and screen save from gems visualisation toolkit viewer, the mr product can generate image headers with non-unique series identifiers, which when exported to some pacs systems, results in the series being associated with the incorrect study on the pacs. the intended study will not have the correct series listed and an unintended study will have it listed instead. patient identifying annotation remains correct. note that related issues have only been seen on pacs systems manufactured by intelerad (intelepacs 4-3-1-p350 and intelebrowser -4-3-1-p350) at this time.
  • Action
    GE Healthcare is providing temporary work around instructions and implementing a software update to correct the issue.

Device

  • Model / Serial
    1.5T Signa HDxt, 3.0T Signa HDxt, Signa Vibrant MR Systems (Magnetic Resonance Imaging Systems)ARTG Number: 108415
  • Manufacturer

Manufacturer