Device Recall Puritan Bennett

  • Model / Serial
    Model number PB980 Ventilator:35B1400143, 35B1400684, 35B1400627, 35B1400708, 35B1400398, 35B1400482,  35B1400491, 35B1400512, 35B1400601, 35B1400121, 35B1400568, 35B1400590, 35B1400595, 35B1400599, 35B1400606, 35B1400106, 35B1400146, 35B1400169, 35B1400226, 35B1400260, 35B1400270, 35B1400263, 35B1400351, 35B1400373, 35B1400400, 35B1400401, 35B1400608, 35B1400089, 35B1400112, 35B1400683, 35B1400685, 35B1400678, 35B1400118, 35B1400114, 35B1400355, 35B1400144, 35B1400300, 35B1400379, 35B1400412, 35B1400472, 35B1400489, 35B1400490, 35B1400492, 35B1400493, 35B1400499, 35B1400500, 35B1400501, 35B1400502, 35B1400507, 35B1400577,  35B1400207, 35B1400211, 35B1400216, 35B1400222, 35B1400225, 35B1400231, 35B1400267, 35B1400312, 35B1400323, 35B1400340, 35B1400383, 35B1400128, 35B1400363, 35B1400459, 35B1400589, 35B1400616,  35B1400637, 35B1400483, 35B1400311, 35B1400447,  35B1400648, 35B1400651, 35B1400654, 35B1400662,  35B1400686, 35B1400692, 35B1400175, 35B1400294, 35B1400436, 35B1400485, 35B1400586, 35B1400644,  35B1400687, 35B1400689, 35B1400694, 35B1400700,  35B1400661, 35B1400676, 35B1400693, 35B1400732,  35B1400734, 35B1400736, 35B1400737, 35B1400742, 35B1400280, 35B1400360, 35B1400471, 35B1400609,  35B1400674, 35B1400688, 35B1400755, 35B1400480,  35B1400675, 35B1400706, 35B1400771, 35B1400667,  35B1400704, 35B1400705, 35B1400727, 35B1400743,  35B1400279, 35B1400285, 35B1400262, 35B1400265, 35B1400259, 35B1400494, 35B1400403, 35B1400779, 35B1400783, 35B1400509, 35B1400724, 35B1400763, 35B1400767, 35B1400566, 35B1400768, 35B1400772,  35B1400791, 35B1400793, 35B1400817, 35B1400525, 35B1400796, 35B1400801, 35B1400809, 35B1400821,  35B1400828, 35B1400333, 35B1400366, 35B1400392,  35B1400440, 35B1400463, 35B1400478, 35B1400308,  35B1400347, 35B1400367, 35B1400396, 35B1400078,  35B1400099, 35B1400130, 35B1400155, 35B1400157, 35B1400158, 35B1400174, 35B1400176, 35B1400178, 35B1400180, 35B1400281, 35B1400315, 35B1400631, 35B1400635, 35B1400625, 35B1400653, 35B1400843, 35B1400847, 35B1400233, 35B1400271, 35B1400337, 35B1400350, 35B1400359, 35B1400368, 35B1400384, 35B1400468, 35B1400486, 35B1400495, 35B1400094, 35B1400236, 35B1400437, 35B1400438, 35B1400479, 35B1400496, 35B1400505, 35B1400506, 35B1400427, 35B1400511, 35B1400325, 35B1400462, 35B1400514, 35B1400536, 35B1400537, 35B1400544, 35B1400066, 35B1400200, 35B1400782, 35B1400829, 35B1400015, 35B1400110, 35B1400516, 35B1400517, 35B1400859, 35B1400013, 35B1400103, 35B1400167, 35B1400124, 35B1400007, 35B1400058, 35B1400083, 35B1400129, 35B1400132, 35B1400006, 35B1400079, 35B1400113, 35B1400116, 35B1400219, 35B1400163, and 35B1400543.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution -- USA, including the states of CA, CO, FL, GA, KY, MA, MN, NC, NY, OH, OK, PA, SC, TN, TX, UT, WI; and, the countries of Canada, Mexico, Saudi Arabia, South Africa, and United Arab Emirates.
  • Product Description
    Covidien Puritan Bennett 980 Ventilator, Rx ONLY. || Suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator for Neonatal (NICU) through Adult patient populations.
  • Manufacturer
  • 1 Event

Manufacturer

  • Manufacturer Address
    Nellcor Puritan Bennett Inc. (dba Covidien LP), 6135 Gunbarrel Ave, Boulder CO 80301-3214
  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    USFDA

12 devices with a similar name

Learn more about the data here

  • Model / Serial
    All Puritan Bennett 980 ventilators
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA (nationwide) and Internationally to Canada.
  • Product Description
    Puritan Bennett 980 Ventilator System, Model No. PB980 Ventilator (980xxxxxxxx), || The Puritan BennettTM 980 Ventilator System is designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator support, delivered invasively or noninvasively, to patients who require the following types of ventilator support: "Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs) "Assist/ Control, SIMV or Spontaneous modes of ventilation.
  • Manufacturer
  • Model / Serial
    PB980 Ventilator (980xxxxxxxx): 980U1ENDIUU, 980U1ENDIUUS, 980U3ENDIUU, 980U3ENDIUUS, 980N1ENDIUU, 980N1ENDIUUS, 980N3ENDIUU, 980N3ENDIUUS
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution-US (nationwide) and the countries of Canada, United Arab Emirates, Argentina, Australia, Bahrain, Brazil, Chile, Colombia, Costa Rica, Egypt, Ireland, Israel, Jordan, Japan, Kuwait, Lebanon, Morocco, Mexico, Panama, Peru, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Turkey, and Uruguay.
  • Product Description
    Puritan Bennett 980 Ventilator System, Universal (with neonatal functionality enabled) and Neonatal Models. || Designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb).
  • Manufacturer
  • Model / Serial
    all versions of software.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution -- US, including but not limited to the territory of Puerto Rico; and Ireland, Jordan, Saudi Arabia, Bahrain, Australia, Japan, Israel, Turkey, Chile, South Africa, Costa Rica, United Arab Emirates, Qatar, Peru, Panama, Singapore, Kuwait, Lebanon, Brazil, Mexico, Colombia, Russia, Uruguay, Slovenia, Morocco, Argentina, Egypt, and Romania.
  • Product Description
    Puritan Bennett 980 Ventilator System, PB980 Ventilator (980xxxxxxxx). || Intended to provide continuous ventilation for pediatric and adult patients who require either invasive ventilation or non-invasive ventilation.
  • Manufacturer
  • Model / Serial
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide Distribution - USA (nationwide) Distribution to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, and the District of Columbia. and to the countries of : Argentina, Chile, Mexico, Panama, Canada, Australia, New Zealand, Austria, Bahrain, Belgium, Bulgaria, Croatia, Czech Republic, Egypt, Estonia, France, Germany, Greece, Hungary, Ireland, Italy, Jordan, Kuwait, Latvia, Lebanon, Lithuania, Morocco, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Israel, Netherlands, Tunisia, Brunei, Indonesia, Malaysia, Myanmar, Philippines, Singapore, Thailand, and Vietnam
  • Product Description
    Rechargeable Li-ion battery used with Puritan Bennett TM 980 Series Ventilator System including Pediatric-Adult Ventilator, Neonatal Ventilator, Universal Ventilator; Battery Product Number: 10086042; UDI: 20884521700861 || The rechargeable lithium-ion batteries are intended to provide a backup power source to the ventilator in case AC power is lost.
  • Manufacturer
  • Model / Serial
    All units ever manufactured. The product may be identified and tracked through serial numbers. This is a tracked device.
  • Product Classification
  • Device Class
    2
  • Implanted device?
    No
  • Distribution
    Worldwide.
  • Product Description
    Model 740 and 760 series Ventilators.
  • Manufacturer
7 more