Reason
Roche has identified that in very rare cases a sample mismatch may occur in specific immunochemistry modules. this issue can only occur if all 5 conditions below are met simultaneously.1. the immunochemistry module is included in its respective system (not standalone); and2. the “module rack buffer setting” ? 1; and3. one rack (rack a) is undergoing sampling and the following rack (rack b) is waiting for measurement; and4. the gripper (t/v carrier) fails to pick up the last vessel on rack a. as a result, the measurement for the sample in the last vessel which failed to be picked up is cancelled and the warning “tips/cup pick up error” (caution level) is issued; and5. sample position 1 of the following rack (rack b) is empty or has no test order for the immunochemistry module.This issue may be detected by 2 alarms occurring within a few minutes of one another: “tip/cup pick up error” (caution level)“abnormal l2-line movement” (s.Stop level).
Action
Roche Diagnostics is requesting customers follow the detailed instructions given in the Customer Letter and to acknowledge receipt of the communication even if they have no stock which is subject to this recall action.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Siemens healthcare has identified a software defect in the stratus cs acute care diagnostics system regarding testpak calibration and quality control (qc) status when a testpak is due to expire within 48 hours. depending on the calibration status, either an above assay range (aar) or an inaccurate value could potentially be reported when running a testpak that is due to expire within 48 hours. the probability of occurrence is extremely unlikely and other factors such as previous results, patient presentation, and other diagnostic testing would initiate clinical questioning and reduce the potential for injury.
Action
Siemens Healthcare is advising users of the following:
1. Do not use TestPaks that are due to expire within 48 hours, OR
2. Check Reagent Status to ensure all TestPaks in use have valid calibration expiry dates beyond the TestPak lot expiry date, AND
3. If the laboratory performs routine QC and if using a TestPak that is due to expire within 48 hours, ensure that scheduled QC is performed before running a test.
Siemens Healthcare is currently developing software updates to address this issue and will be providing new information as it becomes available.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Abbott has identified that the drug fulvestrant (faslodex) may interfere with the architect estradiol assay (ln 7k72) leading to falsely elevated estradiol results. in worst case scenarios, falsely elevated estradiol results may lead to incorrect menopausal state assessment and inappropriate therapy.
Action
Abbott is advising users that patients undergoing Fulvestrant therapy should not be tested with the Architect Estradiol assay. Users are advised to review this letter with the laboratory Medical Director. This action has been closed-out on 20/02/2017.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Roche diagnostics has tested for fulvestrant interference / cross-reactivity with the elecsys estradiol assay; this investigation has shown that fulvestrant could potentially affect patient sample results. the reported issue may lead to falsely elevated results of estradiol. subsequently, misinterpretation of the hormone status could occur and the use of fulvestrant may be altered. in addition, the efficiency of anti-estrogen treatment might be underestimated.
Action
Roche is advising customers that Fulvestrant will increase the apparent concentration of estradiol in women being treated with this drug. A review of estradiol assay results in patients taking fulvestrant is recommended. A disclaimer will be added to the “Interference – Limitations” section in the Method Sheet of the Elecsys Estradiol III assay: This action has been closed-out on 30/01/2017.
“Due to the risk of cross reactivity, this assay should not be used when monitoring estradiol levels in patients being treated with Fulvestrant.”
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
A vitros system software timing anomaly has been identified that could cause two different sample metering scenarios that may lead to erroneous results.Scenario 1: the vitros system could aspirate sample from an unintended sample container causing assay result(s) obtained from that sample to be incorrectly associated with the intended sample.Scenario 2: a sample could be aspirated from a sample container (sample a) and be dispensed into an unintended container (sample b) to be contaminated and diluted by sample a.There is the potential for false negative and false positive results which may lead to an inappropriate diagnosis and patient management leading to serious patient injury.
Action
Ortho is providing users with interim instructions to follow to decrease the probability of the issues occurring, and can review the affected systems to determine if the errors have occurred. It is recommended to discuss any concerns regarding previously reported results with the Laboratory Medical Director to determine the appropriate course of action. Ortho will be providing users with a software update (software version 3.2.3) as a permanent correction. This can be automatically downloaded for systems connected to e-Connect or software kits provided for systems not connected to e-Connect. This action has been closed-out on 20/02/2017.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
There is a potential for an increase in u90-382 or 6lu condition codes (wash errors) when using vitros calibrator kit 9, lot 0954.The increase in u90-382 and 6lu condition codes primarily occurs when using vitros dgxn and phyt slides. vitros acet and crbm slides are not affected. when the u90-382 or 6lu condition code occurs, the vitros system suppresses the result and no result is reported. all numerical results obtained are valid providing that no wash errors occurred and quality control results were within acceptable limits.
Action
OCD is advising users to discontinue use and discard all remaining inventory of the affect lot. OCD will replace the affected units or provide a suitable credit. The affected lot may still be used until replacement lots are available. This action has been closed-out on 06/09/2016.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Siemens is informing customers of potential issue when both ports (serial and ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port. this could cause the message received by the lis (data management system) to include either: duplicate data, missing data and data from a different patient. the risk to health is limited to the extremely unlikely situation in which an analyte is switched between two patients. a transposed sodium, potassium, glucose, lactate, carboxyhemoglobin, methemoglobin, or neonatal bilirubin result may lead to a potential risk to health when the true value is critical.In all cases, the probability of occurrence is extremely unlikely due to the low frequency of the software issue and correlation with patient clinical presentation, history, and concurrent laboratory testing.
Action
Siemens is providing end users with instructions on how to disable Dual Port LIS transfer. The resolution of this issue will be addressed in the next software update.
This action has been closed-out on 01/03/2018.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Hemocue has become aware that the hemocue glucose 201 microcuvette (lot number110718) may, at the end of their shelf-life, yield measurements higher than expected. the higher glucose values may have led to unnecessary insulin treatment or lack of glucose treatment.
Action
The microcuvettes of this specific lot number have now passed their expiry date and should have been used, or be discarded. It is recommended that the need for repeat testing be determined by the laboratory or clinical director. This action has been clsoed-out on 22/08/2016.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Siemens has identified that immulite/immulite 1000 cortisol kit demonstrate a positive bias with patient samples in the range of approximately 13% to 40%. the positive bias was also observed in quality control materials and may result in values outside the established ranges.. the observed positive bias in cortisol values may potentially lead to additional investigation of adrenal status.
Action
Siemens are advising users to discontinue use and discard any remaining kits. The issue has been resolved with IMMULITE/IMMULITE 1000 Cortisol kit lots 384 and above (supplied since 1 August, 2014). A look back of previously generated results is at the discretion of laboratory. This action has been closed-out on 12/08/2016.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Siemens healthcare diagnostics is conducting an urgent recall for product correction for the advia centaur and advia centaur xp immunoassay systems, regarding manually loaded sample racks that are improperly loaded or pushed too far on the sample entry queue during normal operation of the advia centaur and advia centaur xp immunoassay systems. pushing the sample racks may cause misreads of sample ids (sid) if the sample rack loading instructions are not followed exactly as described in the advia centaur or advia centaur xp operator’s guide.If a sample rack is manually pushed with a slight forward nudge to the entry position without placing it on the rack positioning guide, the barcoded tube sid can be assigned to different positions in the rack. this may lead to misreported results.
Action
Siemens Healthcare Diagnostics is advising the customers that when loading a sample rack, the user must ensure that the sample rack notch rests on the raised area of the sample entry queue and the system automatically moves the rack to the loading position. The raised guide ensures the sample rack is positioned correctly and will accurately scan each barcode as it enters the inprocess queue. Users are further advised not to push the sample racks into the system.
During the next service visit, Siemens Healthcare Diagnostics service engineers will affix a label stating "Do not push sample rack" onto the system to reinforce the proper loading technique of the sample rack. This action has been closed-out on 5/08/2016.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
An abbott investigation has identified that the ict serum calibrator lots above may generate lower than expected potassium qc and patient results. a potential cause of this issue is increased ammonia content over the shelf life of the calibrator. elevated ammonia concentration may impact the potassium calibration and potentially cause a calibration curve to generate the lower results. patient results may have shifted down by 1-3% beyond the acceptable performance of the assay.
Action
Abbott is advising their customers that if the calibration curve for Potassium is verified with at least two levels of controls according to established quality control requirements for the laboratory, continue to use the lot until new lots are received through the normal ordering processes. If QC results for Potassium are not within established quality control requirements, uses are advised to recalibrate using a different calibrator lot. This action has been closed-out on 03/02/2017.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Investigation at abbott molecular identified that some of the 5ml reaction vessels (rv) contained in lots 56944001, 57026001, 57059001 and 57673001 may have insufficient rim. an rv with an insufficient rim may not be held securely in the 1ml subsystem carrier. this may cause the rv to be seated incorrectly in the carrier or drop through it. it is estimated that less than 1% of rvs within these lots may be defective. these lots have been in distribution since july 2012.
Action
End users are advised to remove them from use any remaining RVs from lots 56944001, 57026001, 57059001, or 57673001 (original vendor lot 2198077). Replacement product will be coordinated on return of the receipt acknowledgement form.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Manufacturer confirms that the solid phase reagent in some of the advia centaur tni-ultra readypacks, lot 084, is darker in appearance, creating a potential for incorrect control and patient results.
Action
Siemens is providing work around instructions to users. Only one TNI-Ultra ReadyPack is to be used on a system at one time. Patient results produced by the assay are acceptable if they follow acceptable calibration with valid quality control results. Kit lots ending in 086 or higher do not require calibration of each ReadyPack, refer IFU for the calibration interval. This action has been closed-out on 11/04/2016.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Beckman coulter has confirmed that some ise units can experience premature ise ref (reference) valve failure. premature valve failure can be detected by the presence of air bubbles in the ise ref solution tubing on the ise unit. if air bubbles are present, accuracy of the ise test results may be affected. affected results may be clinically significant (high or low) and may impact patient diagnosis and/or treatment.
Action
Beckman Coulter is providing users with instructions on how to inspect tubing for the presence of air bubbles. Where air bubbles are present, users are to contact Beckman Coulter for a replacement valve. Beckman Coulter is recommending consulting the laboratory director to determine the impact of this issue and if a retrospective review of prior results is necessary.
As a preventative measure, Beckman Coulter is proving work around instructions for users to follow until a replacement ISE valve is implemented.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
A software bug has been detected in the cobas integra 800 software version 9864.C2 which affects the execution of routine maintenance of the ise module. the issue is caused by the ise counters in the software being reset to zero whenever the ise rack is removed from the analyser. lf ise service actions and manual maintenance items are not executed at their recommended intervals, there is a potential risk of increased occurrence of calibration flags (slope, sol i dev etc), as well as incorrect and unflagged ise results.
Action
Roche is currently working on a software fix expected to be implemented by the end of the year. Until the software is available, Roche recommends that operators manually track all ISE counters and to manually request the corresponding Service Action in the Service Software.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Radiometer medical has become aware that some pco2 membranes can cause a negative bias, on both blood and quality control measurement results.The bias can be seen after the replacement of the membrane.
Action
Radiometer Medical are providing additional quality control instructions to ensure the accuracy of pCO2 results. This action has been closed-out on 10/02/2016.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Beckman coulter has confirmed customer reports of the affected lot failing calibration with error condition ocr low. the calibration failures appear to be related to improper shipping or storage conditions.
Action
Beckman Coulter is requesting end users to discontinue use of the affected lot. Re-assessment of past results is not required as the LACT results produced after acceptable QC recoveries on each reagent cartridge are accurate.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Siemens previously communicated an issue with low bias on patient results for dimension tacrolimus flex reagent cartridge lot fa3316 (urgent medical device recall dc 13-02a and tga ref rc-2013-rn-00050-1). siemens will be implementing linking of specific dimension tacr calibrator lots with specific dimension tacr flex lots to ensure accurate recovery and continued traceability of patient results to the lcms reference method. due to the requirement for linked calibrators, lots 2ed072 and 2kd084 can no longer be used.
Action
Siemens is requesting customers to immediately discard any remaining Dimension TACR (DF107) Flex lot 2ED072 and 2KD084. To ensure the correct linkage between calibrator and reagent lots, customers are requested to contact their Siemens customer service to order lot linked combinations.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
This product contains detectable levels of hcv rna which originated from the non-specific human igm component used in the mouse monoclonal antibody conjugation process for this product.
Action
Bio-Rad is asking users to destroy the affected stock using appropriate disposal methods.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Issue 1: loose drive support cap – if the loose drive support cap is pressed on, it could result in the unintended delivery of additional insulin.Issue 2: sensor graph timeout (only affects paradigm veo (mmt-554 and mmt-754)) – this issue could prevent the auto-resume of basaldelivery 2-hours after a low glucose suspend event, which can result in elevated blood glucose values.Issue 3: water damage– this issue may result in a pump alarm or cause the buttons to stop working.
Action
Issue 1: Medtronic is asking consumers to regularly examine the pump especially after a shock or drop on a hard surface. If the drive support cap is sticking out, customers are asked to not press on this portion of the pump and to discontinue the use of the pump. Customers will need to manage their diabetes as per the back-up plan provided to them by their healthcare professional. Also, contact Medtronic 24 hour Helpline on 1800 777 808 (option 1).
Issue 2: Medtronic is providing work around instructions to prevent this issue from occurring.
Issue 3: Medtronic is asking customers to follow the user guide to prevent this issue from occurring.
Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
Extra notes in the data
Reason
Ortho clinical diagnostics has received customer complaints of intermittent imprecision, outliers, or shifts in accuracy and control fluid values outside of expected intervals when using vitros chemistry products phbr slides. an investigation confirmed that positively biased values for patient samples could be obtained that may not be detected by quality control fluids.
Action
End users are requested to discard all lots of VITROS PHBR Slides. It is recommended that any concerns with previously reported patient results be discussed with the laboratory director and requesting physician.
Irish data is current through April 2019. All of the data comes from the Health Products Regulatory Authority (Ireland), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Ireland.
Extra notes in the data
Advice regarding changes/updates made to instructions for use. 3rd Party Publications
Irish data is current through April 2019. All of the data comes from the Health Products Regulatory Authority (Ireland), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Ireland.
Extra notes in the data
Advice regarding a device removal. 3rd Party Publications
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Extra notes in the data
in vitro diagnostic device for quantitative determination of blood glucose level. - Product Code CGA
Reason
Incorrect configurations settings in software. as a result, certain error codes do not trigger appropriately. these error codes are: e01, e02 and e05 which are all related to the optronic performance of the analyzer. possible incorrect results.
Action
Hemocue AB sent Customer Notification "Notice of Medical Device Correction on HemoCue Glucose 201 DM and HemoCue Hb 201 DM Analyzers" dated October 7, 2010, to customers notifying them of incorrect settings on error codes.
The corrective action consists of an update of the configuration settings on the the affected HemoCue Glucose 201 DM and Hb 201 DM analyzers that are in the field. The affected units in the field will be corrected in accordance with a rework procedure that has been established by HemoCue AB.
U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
Extra notes in the data
Real time nucleic acid amplification system - Product Code OOI
Reason
If the bar code is printed from the batch results screen, a wrong batch id may be printed, while the correct batch id is still assigned to the batch in the data repository of the mplc 2.0 software. this may result in sample mis identification. if printed from the ordering screens, the bar code is printed with the correct batch id.
Action
The firm, Roche, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 04, 2010 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to print batch ID barcode labels from the ordering or stage set up screens only; print Cooling Block barcode labels from the Post Elution Edit screen only; complete and return the attached UMDC Faxback Form for MagNA Pure LC 2.0 Instrument [UMDC 10-228] via fax to 1-877-551-7152 and file this correction notice for future reference. The firm plans to issue a software update in first quarter 2011.
If you have questions about the information contained in this UMDC, for Clinical customers, please contact Roche Molecular Diagnostics Technical Service at 1-800-526-1247; for Research customers, please contact Roche Applied Sciences Technical Service at 1-800-262-4911.