Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It was stated by the manufacturer that at the time of performing the calibration of the equipment, the system may fail due to a bias in the chloride parameter, leading to errors in the results of the diagnosis and potentially adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It is affirmed by the manufacturer that the welchallyn parent company will end with the manufacture and sale of the mentioned references after december 2014, in particular propaq cs with the option of co2 was suspended since january 24, 2014, however, informs that it will continue to provide support and technical service for up to five (5) more years or until the spare parts and accessories are exhausted.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It has been determined that the reconstitution time of 30 minutes is insufficient to reconstitute the lots of thyroglobulin calibrator.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It informs that the mixing matrices can cause software errors and the results obtained, leading to potentially adverse events occurring on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It is reported that the pet reconstruction server (prs) can be blocked after the acquisition of ct in low doses and may not allow the acquisition of pet, which would result in incomplete cases, therefore potential for erroneous diagnoses and adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It is reported that the manufacturer issues the withdrawal of 113 units due to a possible nonconformity in the manufacturing process, leading to the generation of possible adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
He mentions that lithium batteries have deformation to increase its dimensions by chemical reactions inside it, which can separate the shell of the pump to prevent the malfunctioning of the biomedical equipment, leading to the occurrence of potentially adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It states that the assembly of the articulated arm used for the assembly of the camera can be detached and cause it to fall, leading to the occurrence of potentially serious adverse events on the patient or the operator.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
He mentions that there is a potential risk associated with the use of delelectrode reference, since the kcl solution can escape slowly from the electrode creating an electrode parallel to the reference, situation that can cause a change in the ph and sodium readings, leading to potential events adverse on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
The manufacturer reports that it has developed a new syringe demonstrates 100μl reference b06040 to be used in the medical devices in question and thus counteract the deterioration of the prematurely damaged elements previously used for this purpose.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It is affirmed that the ventilators of the models mentioned commercialized between december 31, 2013 and january 30, 2014, contain a potentially defective ferrite component that could affect the operation of the biomedical equipment, causing a failure in the mechanical ventilation and in the functionality of the alarm that can be reduce only to indicate ventilatory insufficiency, leading to the potential occurrence of adverse events on the patient causing serious deterioration in their health condition or even death.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
The manufacturer informs that there have been differences between the real physical positions and the reading made by the equipment of the table and the table, due to the fact that the connection lines have a very high resistance, providing differences in the reference voltages to the spatial deposition potentiometers, leading to that potential adverse events are presented to the patient or the user.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
The manufacturer reports that the articulation of the mechanical system between the scissor arm and the support could fail prematurely, due to a manufacturing problem for the models mentioned and that were manufactured between may 2008 and april 2010, leading to the occurrence of potentially adverse events. the patient or the user.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
The manufacturer informs that the filters for the ecg signal selected by the users can be affected by an additional high-pass filter of 0.5 hz, which would generate changes in the ecg wave morphology and values on the screen of stincorrect, leading to there are potentially adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It informs that the detector of radial movement of the biomedical equipment can come into contact with the patient due to involuntary movement during its configuration or at the end of the study, leading to the occurrence of potentially adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
Affirms that the health agency of france (ansm) found that the sterilization processes used in the manufacture of this device have not been fully validated in accordance with international regulations for which you must suspend the marketing and use of them until the investigation determines that they are safe for its use or sedecida the definitive withdrawal of the product from the market, additionally states that have not yet been identified health risks associated with the implanted prostheses.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
The manufacturer informs that the software could visualize and print the name of a patient about a study that does not correspond, this error affirms it is related to the gluing impression of microsoft word, leading to the potential occurrence of adverse events on the patient.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It presents a negative bias of 50% in the results of the patients.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
The manufacturer reports that in systems with software versions 6.0, 6.1 and 6.2, the age calculation displayed on the screen may not be correct (± 2 weeks with respect to actual age), although this calculation does not affect the age of the patient initially admitted. to the system, could lead to medical decisions that are based on the system being affected.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
They present problems with optical degradation, which generates delays in obtaining results for erythroblast, leukocyte count and reticulocytes.
Colombian data is current through June 2019. All of the data comes from the National Institute of Drugs and Food Surveillance (INVIMA), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Colombia.
Extra notes in the data
Reason
It is reported that they may experience a temporary change in their correct functioning when exposed to currents emitted by sources of electrocautery, this change may last thirty (30) seconds or more after the exposure is determined, which may lead to potential adverse events on the patient.