translation missing: en.activerecord.attributes.event.titles.recall_field_safety_notice

According to Saudi Food & Drug Authority, this translation missing: en.activerecord.attributes.event.types.recall_field_safety_notice involved a device in Saudi Arabia that was produced by 3M Health Care Ltd.

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Recall / Field Safety Notice
  • Event ID
    mdprc 103 01 18 000
  • Event Initiated Date
    2018-01-31
  • Event Country
  • Event Source
    SFDA
  • Event Source URL
  • Notes / Alerts
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • Extra notes in the data
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • Reason
    3m confirmed that that the affected blankets are more susceptible to unintended occlusion of airflow resulting in partial or limited inflation during use. if blankets are only partially inflated during use in surgery, there is potential for incomplete warming therapy to be given to a patient.

Device

  • Model / Serial
    1) 3MTM Bair Hugger™ Full Access Underbody Warming Blanket 63500 Lot Nos.: R10359, R10360, R10361, R10362, R10363, R10364, R10365, R10366 2) 3MTM Bair Hugger™ Full Access Underbody Warming Blanket 63700 Lot Nos.: FR10361
  • Product Description
    Warming Units, Patient, Forced-Air
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturers representative
    3M company
  • Source
    SFDA