Field Safety Notices about ARCHITECT Anti-HCV Reagents

According to Health Products Regulatory Authority, this field safety notices involved a device in Ireland that was produced by Abbott GmbH & Co. KG..

What is this?

Field safety notices are communications sent out by medical device manufacturers or their representatives in relation to actions that they may be taking in relation to their product that is on the market. These are mainly for health workers, but also for users. They can include recalls and alerts.

Learn more about the data here
  • Type of Event
    Field Safety Notice
  • Event Number
    SN2010(02)
  • Date
    2010-04-15
  • Event Country
  • Event Source
    HPRA
  • Event Source URL
  • Notes / Alerts
    "Irish data is current through April 2019. All of the data comes from the Health Products Regulatory Authority (Ireland), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Ireland."
  • Extra notes in the data
  • Reason
    Abbott has determined that s/co values generated with architect anti-hcv (ln 6c37) may decrease when 25% or less of test volume is remaining leading to:  • abbott positive control values shift down and/or out of range low, • non-abbott positive control values shift down and/or out of range low, and/or • a decrease in patient result values.
  • Action
    Abbott has issued a Field Safety Notice (FSN) (March 2010) to all Irish customers using the ARCHITECT Anti-HCV assay recommending customers mix the ARCHITECT Anti-HCV assay diluent (LN 6C37J, green bottle label) prior to testing each day of use.

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    HPRA