Recall of SYSTEM 83+

According to Health Canada (via FOI), this recall involved a device in Canada that was produced by CUSTOM ULTRASONICS INC..

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    23858
  • Event Risk Class
    I
  • Event Initiated Date
    2015-12-04
  • Event Country
  • Event Source
    HC
  • Notes / Alerts
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • Extra notes in the data
  • Reason
    Custom ultrasonics inc. failed to complete validations of the olympus evis exera ii tjf q180v duodenoscope processed in the system 83 plus and the system 83 plus 9 washer disinfectors and failed to complete validation of various high level disinfectants used with the system 83 plus 2 and system 83 plus 9 washer disinfectors.

Device

  • Model / Serial
    Model Catalog: SYSTEM 83 PLUS 2 (Lot serial: >10 CONTACT MFR); Model Catalog: SYSTEM 83 PLUS 9 (Lot serial: >10 CONTACT MFR)
  • Product Description
    SYSTEM 83+
  • Manufacturer